Fundamentals of Analytical Toxicology. Robin Whelpton
Чтение книги онлайн.

Читать онлайн книгу Fundamentals of Analytical Toxicology - Robin Whelpton страница 20

Название: Fundamentals of Analytical Toxicology

Автор: Robin Whelpton

Издательство: John Wiley & Sons Limited

Жанр: Биология

Серия:

isbn: 9781119122371

isbn:

СКАЧАТЬ Fischer The absolute stereochemistry in the Fischer notation gives the absolute spatial arrangement by reference to D-glyceraldehyde. The letters D or L are used (not to be confused with lower case d or l). The Fischer convention is still used for carbohydrates and amino acids. The original choice of D-glyceraldehyde was arbitrary, but was proved correct by X-ray crystallography Cahn–Ingold–Prelog (CIP) The CIP system is the definitive method of assigning absolute configuration. The letters R and S indicate spatial arrangements as follows:Assign values to the substituent groups by highest atomic numberaPoint the lowest value awayIf the remainder go from high to low clockwise then R (rectus)If the remainder go from high to low anticlockwise then S (sinister)In case of a tie go to the next atoms along The Fischer convention cannot be simply converted to the CIP system, i.e. R does not always equate to D. All naturally occurring α-amino acids in mammalian proteins are L-. Using the CIP system, cysteine and cystine are S- the others, without sulfur, are R-

      aThe rules are in fact more detailed: Highest atomic number > highest atomic mass > cis- prior to trans- > like pairs (RR) or (SS) prior to unlike > lone pairs are considered an atom of atomic number 0

      Quantitative methods must have good precision (reproducibility) and accuracy (the results must reflect the true concentration of the analyte). Selectivity (freedom from interference, specificity) is important when a single species is to be measured, but broad specificity may be useful when screening for the presence of a particular class of compounds as discussed above. The recovery of the analyte, i.e. the proportion of the compound of interest that is recovered from the sample matrix during an extraction, for example, is important if sensitivity is limiting, but need not be an issue if the LLoD, accuracy, and precision of the assay are acceptable.

Absolute stereochemistry The absolute spatial configuration of atoms of a molecule
Chiral Hand-like, i.e. left- and right-handed mirror images
Enantiomer One mirror image from of a pair of non-superimposable optically active compounds
Epimers Optically active molecules with more than two chiral centres differing at only one chiral centre
Epimerization Partial racemization of one chiral centre in a molecule with two or more chiral centres
Diastereomers Stereoisomers that are not mirror images of each other
Inversion Conversion of one enantiomer to the other
Meso Optically inactive isomer in which the activity of chiral centres are balanced
Racemate Equimolar mixture of both enantiomers of an optically active compound
Racemization Conversion of a single enantiomer to a racemate

      Ideally, whatever the methodology employed, quantitative assay calibration should be by analysis of standard solutions of each analyte (normally 6–8 concentrations across the calibration range) prepared in the same matrix and analyzed as a batch along with the test samples. A graph of response against analyte concentration should be prepared and used to calculate the analyte concentration in the sample (so-called ‘external standard’ method). Use of a full calibration sequence may not be possible in emergency toxicology, for example, and in such circumstances single-point calibration is often acceptable if properly validated (Section 3.2.4.10).

      Any quantitative analysis is a measurement and, in common with all measurements, has associated errors, both random and systematic. In chromatographic and other separation methods the ‘internal standard’ method is often used to reduce the impact of systematic errors such as variations in injection volume or evaporation of extraction solvent during the analysis. Thus, a known amount of a second compound that behaves similarly to the analyte during the analysis, but elutes at a different place on the chromatogram or is otherwise detected independently of the analyte (the internal standard, ISTD) is added at an appropriate stage in the analysis. Subsequently, the detector response of the analyte relative to the response of the ISTD is plotted against analyte concentration when constructing a calibration graph (Section 3.3). In the case of MS methods, stable isotope-labelled analogues of the analyte should be used if available. The use of a marketed medicament should be avoided otherwise a significant bias may result if the patient has taken this drug.

       1.3.4 Quality control and quality assessment

      Once an analytical method has been validated and implemented it is important to be able to show that the method continues to perform as intended. In qualitative work, known positive and negative specimens should normally be analyzed in the same analytical sequence as the test sample. A negative control (‘blank’) helps to ensure that false positives (owing to, for example, contaminated reagents or glassware, or carry-over from a previous analysis) are not obtained. Equally, inclusion of a true positive serves to check that any reagents have been prepared properly and have remained stable.

Term Notes
Accuracy The difference between the measured value and the accepted (‘true’) value
Calibration range The range of concentrations between the highest and lowest calibration standards. This should encompass the range of concentrations found in the test samples
Carry over Signal from the last sample analysis enhancing response in the next analysis
Carry under Signal from the last sample analysis reducing response in the next analysis
Coefficient of variation (CV) An alternative term for relative standard deviation (RSD)
‘Cut-off’ An analyte concentration above or below which positive results are not reported (see Section 18.1.2)
Higher or upper limit of quantification (HLoQ/ULoQ) The highest concentration that can be quantified. Not always quoted, but important in assays with a clear upper limit to a calibration graph, for example immunoassays and fluorescence assays
Internal quality control (IQC) A sample of known analyte concentration that is analyzed as if it were a real sample
Internal standard (ISTD) A second compound, not СКАЧАТЬ