Название: Continuous Emission Monitoring
Автор: James A. Jahnke
Издательство: John Wiley & Sons Limited
Жанр: Биология
isbn: 9781119434023
isbn:
The Electric Power Research Institute developed an algorithm for in‐stack and external dilution system corrections in their “CEMS Analyzer Bias and Linearity Effects (CABLE) study” (Berry 2000). The study was conducted to improve the accuracy of CEM systems used in the acid rain program and provided recommendations for the implementation of a correction algorithm based on Equation 3‐9. In this algorithm, the temperature dependence, f(T), was expressed empirically as f
Dilution Extractive Systems: Wet or Dry Measurement?
A question that frequently arises when considering the use of a dilution system is whether the emission values are given on a wet basis or a dry basis. Dilution systems that do not dry the diluted sample gas give concentration values (ppm) on a wet basis. If there is no moisture removal, there will still be moisture in the diluted sample. When the diluted concentration (measured with ambient air level analyzers) is scaled back up by multiplying it by the dilution ratio, the moisture content is scaled back up also. The concentration of the pollutant gas in the original wet sample is obtained.
Consider an example of 1 ml of flue gas that contains 0.1 ml of water vapor (10% H2O), which is diluted with 99 ml of dry air (Figure 3‐27). In the diluted sample, the percentage of water vapor is reduced to about 0.1% of the total volume. Essentially, the analyzer is measuring SO2 in a background gas containing 99.9% dry gas and 0.1% water vapor. Therefore, the flue gas SO2 concentration obtained is given on a wet basis. When the diluted concentration is scaled back up by multiplying it by the dilution ratio, the concentration of SO2 in the original wet sample is obtained.
Figure 3‐27 Example of a wet flue gas being diluted with dry air.
Consider again from Figure 3‐27 that the diluted SO2 concentration is measured to be 3 ppm. This means that for every 100 ml of gas analyzed, the volume of SO2 equals 3 × 10−6 × 100 = 3 × 10−4 ml. However, because no SO2 is in the dry air used for dilution, this must be the volume of SO2 gas contained in the 1 ml of flue gas drawn into the dilution system. A volume of 3 × 10−4 ml in 1 ml is equivalent to a flue gas concentration of 3 × 10−4 × 106 = 300 ppm. Because, in this example, the 1 ml of flue gas sample also contains 0.1 ml of water vapor, the 300 ppm value is actually a wet‐basis value.
To report a dry‐basis SO2 emission concentration, the moisture content must be known. It can be calculated by using Equation 3‐11:
where
cd = dry‐basis concentration value (ppm or percent)
cw = wet‐basis concentration value (ppm or percent)
Bws = the moisture fraction of the flue gas
A value for Bws can be obtained by either installing some type of moisture monitor, using a value obtained by manual stack testing, or by estimating a value based on process parameters. Using a moisture analyzer is the most straightforward approach; however, this adds another analyzer to the CEM system. A common technique is to measure oxygen on both a wet and a dry basis and using the results to compute Bws. Also, various calculation approaches have been proposed (Aldina 1985; McGowan 1976) that involve the manipulation of combustion source F factors. But frequently, a moisture value obtained from manual stack test measurements is set as a constant factor to make the appropriate corrections. This approach assumes that variation in the value will be small under normal source operating conditions.
It is not necessary to convert dilution measurements into a dry basis if emission measurements are to be reported on an actual or wet basis. If the emissions are to be reported in terms of a pollutant mass rate (pollutant mass/unit time) as in the acid rain program, the wet basis measurement can be used directly. Also, if emissions are reported in terms of ng/joule or lbs/million Btu, a moisture correction is not needed if CO2 is monitored along with the pollutant. This is due to the form of the F factor equation used to calculate emission rate in ng/Joule or lbs/million Btu as discussed earlier (see Equation 3‐10). When CO2 is monitored as a diluent and the Fc factor equation is used, moisture content cancels out in the ratio between the pollutant concentration and CO2 concentration. This is also discussed in EPA Test Method 19 (U.S. EPA 2020b).
Lastly, one should be aware of inadequacies associated with the use of dry calibration gases to check the performance of wet systems. Cylinder gases used for calibration or audit purposes are dry gases, containing no water. The molecular weight of a dry gas used to establish the calibration of a dilution system will be different than the molecular weight of a flue gas containing percent levels of moisture. As discussed earlier, this will affect the dilution ratio and introduce bias into the measurement. Also, some analyzers are subject to interference by water vapor. Here, a dry calibration gas may not truly establish the calibration function for the analyzer and the wet flue gas measurements will be in error. Additionally, water can adsorb onto sample tubing and internal instrument surfaces and subsequently reduce the adsorption of other gases such as volatile organic compounds (Peeler et al. 1997). Daily calibration with dry gases can disrupt this effect and cause some disequilibrium in the measurements for a period of time after calibration.
SAMPLING INTERFACE/MONITOR CALIBRATION
The entire extractive system sampling system and gas analysis system must be capable of being calibrated as a unit. In the design of the system, calibration gases should be able to be injected as close as possible at the probe – a recommendation made in fact by the EPA in its Appendix F quality assurance requirements and a requirement for cylinder gas audits and linearity checks. This is necessary to check for leaks or other operational problems in the system. In calibrating source‐level‐extractive systems, the analyzer should be calibrated at the same gas flow rate, pressure, temperature, and operating procedures that are used in monitoring the stack gas. In calibrating dilution systems, the calibration gas must enter the system before the critical orifice. Flooding the coarse filter or dilution probe with calibration gas СКАЧАТЬ