Continuous Emission Monitoring. James A. Jahnke
Чтение книги онлайн.

Читать онлайн книгу Continuous Emission Monitoring - James A. Jahnke страница 50

Название: Continuous Emission Monitoring

Автор: James A. Jahnke

Издательство: John Wiley & Sons Limited

Жанр: Биология

Серия:

isbn: 9781119434023

isbn:

СКАЧАТЬ temperature changes. The system can also provide for changes in gas density as the flue gas moisture or oxygen concentrations vary.

      The Electric Power Research Institute developed an algorithm for in‐stack and external dilution system corrections in their “CEMS Analyzer Bias and Linearity Effects (CABLE) study” (Berry 2000). The study was conducted to improve the accuracy of CEM systems used in the acid rain program and provided recommendations for the implementation of a correction algorithm based on Equation 3‐9. In this algorithm, the temperature dependence, f(T), was expressed empirically as fleft-parenthesis upper T right-parenthesis equals RootIndex x StartRoot upper T EndRoot slash RootIndex x StartRoot upper T Subscript o Baseline EndRoot when the study data were found not to follow the theoretical relation f left-parenthesis upper T right-parenthesis equals StartRoot upper T EndRoot slash StartRoot upper T EndRoot Subscript o. Here, x, is determined for each individual dilution system. Step‐by‐step procedures and a logic diagram for implementation of the dilution ratio correction routine can be found in an EPRI report of the study (Berry 1999).

      Dilution Extractive Systems: Wet or Dry Measurement?

      A question that frequently arises when considering the use of a dilution system is whether the emission values are given on a wet basis or a dry basis. Dilution systems that do not dry the diluted sample gas give concentration values (ppm) on a wet basis. If there is no moisture removal, there will still be moisture in the diluted sample. When the diluted concentration (measured with ambient air level analyzers) is scaled back up by multiplying it by the dilution ratio, the moisture content is scaled back up also. The concentration of the pollutant gas in the original wet sample is obtained.

Schematic illustration of example of a wet flue gas being diluted with dry air.

      Consider again from Figure 3‐27 that the diluted SO2 concentration is measured to be 3 ppm. This means that for every 100 ml of gas analyzed, the volume of SO2 equals 3 × 10−6 × 100 = 3 × 10−4 ml. However, because no SO2 is in the dry air used for dilution, this must be the volume of SO2 gas contained in the 1 ml of flue gas drawn into the dilution system. A volume of 3 × 10−4 ml in 1 ml is equivalent to a flue gas concentration of 3 × 10−4 × 106 = 300 ppm. Because, in this example, the 1 ml of flue gas sample also contains 0.1 ml of water vapor, the 300 ppm value is actually a wet‐basis value.

      where

       cd = dry‐basis concentration value (ppm or percent)

       cw = wet‐basis concentration value (ppm or percent)

       Bws = the moisture fraction of the flue gas

      A value for Bws can be obtained by either installing some type of moisture monitor, using a value obtained by manual stack testing, or by estimating a value based on process parameters. Using a moisture analyzer is the most straightforward approach; however, this adds another analyzer to the CEM system. A common technique is to measure oxygen on both a wet and a dry basis and using the results to compute Bws. Also, various calculation approaches have been proposed (Aldina 1985; McGowan 1976) that involve the manipulation of combustion source F factors. But frequently, a moisture value obtained from manual stack test measurements is set as a constant factor to make the appropriate corrections. This approach assumes that variation in the value will be small under normal source operating conditions.

      The entire extractive system sampling system and gas analysis system must be capable of being calibrated as a unit. In the design of the system, calibration gases should be able to be injected as close as possible at the probe – a recommendation made in fact by the EPA in its Appendix F quality assurance requirements and a requirement for cylinder gas audits and linearity checks. This is necessary to check for leaks or other operational problems in the system. In calibrating source‐level‐extractive systems, the analyzer should be calibrated at the same gas flow rate, pressure, temperature, and operating procedures that are used in monitoring the stack gas. In calibrating dilution systems, the calibration gas must enter the system before the critical orifice. Flooding the coarse filter or dilution probe with calibration gas СКАЧАТЬ