СКАЧАТЬ and compressor‐operated gas coolers are most commonly used for this purpose in extractive systems. Coolers should be correctly sized for the sample gas flow rates and stack‐gas moisture content; also, coolers must remove condensed water rapidly from the gas stream to minimize contact with dried gas.

       Thermoelectric Coolers.

      Hot flue gases can be cooled in thermoelectric Peltier coolers, which take advantage of the Peltier effect to chill the sample gas. The Peltier effect occurs when two dissimilar metals or semiconductors are joined in a loop and a voltage source generates a current through the loop. Because of the differing electron distributions in the dissimilar materials, a small voltage difference exits across their junction. The junction will heat up where the voltage difference opposes the voltage difference of the battery. At the other junction, thermal energy is absorbed from the surroundings and converted into electrical energy to balance the electron flow. This absorption, of course, reduces the temperature of the surroundings.

      In the impinger, the flue gas flows through the central tube, which is surrounded by a vacuum jacket. The flue gas remains above the dew point until it reaches the bottom where it is cooled rapidly. Water vapor condenses and is removed at the bottom of the impinger by typically using a peristaltic pump. The gas is cooled further by the cool impinger wall as it moves up to the impinger outlet. In a unique feature of the design, the gas is reheated by the unjacketed part of the central tube before exiting to the analyzer.

      The effectiveness of this type of cooler is dependent upon the surface area and length of the impinger, gas flow rate, materials of construction, temperature of the ambient air, and the cold side temperature. The sizing of a Peltier cooler/impinger condensation system is dependent upon the demands of the total sampling system, so care must be taken in their proper application. The chillers normally operate to output a gas having a dew point of 4 °C (0.5% moisture concentration), but subzero coolers are available down to −25 °C to reduce the residual moisture concentration and to prevent the formation of acid gases such as SO3.

Schematic illustration of impinger (laminar heat exchanger) used with a Peltier cooling system.

       Compressor Gas Coolers.

      Another condenser can be added to further reduce the water content, but a more efficient technique is to place the sample pump after the first coil and transport the gas from the chiller, under pressure, to a second chiller as shown in Figure 3‐11. A gas under pressure will condense more readily than a gas under vacuum. Since it is more difficult for water molecules to escape the liquid surface to vaporize when the gas is under pressure, this pressurization will, in effect, reduce the moisture content to a level lower than the content under atmospheric pressure.

Schematic illustration of refrigerated condenser moisture removal system with a secondary chiller under positive pressure.

       NafionTM Dryers.

      Another method of removing water vapor from a flue gas sample is to use a material called “Nafion.” Nafion is a copolymer of tetrafluoroethylene‐perfluoro 3, 6‐dioxa‐4‐methyl‐7‐octene sulfonic acid (Permapure 2019). The sulfonic acid group of the copolymer has high water‐of‐hydration, absorbing up to 13 molecules of water. Water vapor will transfer across the material by absorption as water‐of‐hydration if the partial pressure of the water vapor is different on each side of the membrane. Water molecules are initially bound on the material surface. However, unhydrated sulfonic acid groups deeper in the Nafion membrane will have higher affinity for water molecules and will absorb them and in turn pass them onto other unhydrated sulfonic acid groups, causing them to migrate into and through the membrane.

      The Nafion dryer offers a number of advantages over refrigerated chillers because no mechanical parts are incorporated into the system, no condensate trap is required, and the question of pollution absorption in condensed water is avoided. However, dryers using small‐diameter Nafion tube bundles can be prone to plugging, either from droplets of condensed material or from particulate matter introduced by improperly filtered samples or by the precipitation of salts. Problems of condensing liquids may be minimized by heating the entrance side of the dryer, whereas particulate plugging and precipitation can be resolved by using larger diameter tubing. When using large‐diameter tubing (1/4 in.), greater СКАЧАТЬ