Название: Biopolymers for Biomedical and Biotechnological Applications
Автор: Группа авторов
Издательство: John Wiley & Sons Limited
Жанр: Химия
isbn: 9783527818303
isbn:
Another key consideration for the implant test for a biopolymer is with the implantation time points. ISO 10993‐6 states: “For absorbable materials, the test period shall be related to the estimated degradation time of the test product at a clinically relevant implantation site. When determining the time points for sample evaluation, an estimation of the degradation time shall be made.” Usually, in practice we try to estimate the absorption profile based on the specific metabolism rate and method of the material and the implant system. After this, we set three time periods: one where we first see degradation (usually between two and four weeks), second when half the sample is degraded, and third when we see a “steady state” in the sample material. A steady state is defined as a point in time where the body is no longer interacting with the material and no additional changes are happening. For example, in vivo implantation tests with a PLLA density scaffold demonstrated fast degradation in the first three weeks, after which the degradation rate progressively decreased [20]. This milestone is reached when the body has either encapsulated or otherwise dealt with the foreign material or when full degradation of the material has occurred.
As mentioned above, an appropriate control is the basis for the acceptance criteria of the test itself, making it an essential component for a relevant and applicable test system. The implantation test is set up so that the evaluation is conducted by comparing the result of the test site histopathology with the control site. Thus, if the chosen control article is a hard piece of metal or plastic that would not induce interaction with the surrounding tissues, then the comparison with the implant site of the biopolymer would probably not be favorable, leading to a higher tissue reactivity and making it look like the test material is non‐biocompatible. However, if an appropriate control is used, then the histopathological comparison of the test and control article sites can be made with confidence, and a correct understanding of the implantation risk of the material can be drawn.
1.5 Conclusion
Biopolymers occupy a unique and advantageous space as a medical device material. Devices made from these naturally occurring or biomimetic substances have the distinct advantage that the material itself is akin to those tissues the device contacts. From a bulk perspective, there is no concern regarding the material as a foreign body. Biopolymers also have environmental and manufacturing advantages as they are often produced not from petroleum derivatives but by living systems.
In contrast to the major advantages presented by biopolymers within the context of biocompatibility, there are a couple of key concerns that must be addressed. The natural origin of these materials does not mean that they are free from manufacturing residuals. Contact with solvents through manufacturing and purification steps can introduce contamination, as can contact with storage and primary packaging materials. Chemical analysis screening for these compounds can be complicated by the complex organic nature of the device material. Additionally, many biopolymers are degradable or resorbable by the body. While this is, in principle, a positive therapeutic effect, it can be difficult to prove that the safety of the device does not change over the degradation lifetime.
The pallet of materials afforded by biopolymers allows an even broader spectrum of medical devices with huge potential to help mankind. The biocompatibility principles discussed in this chapter can be applied to biopolymers to address concerns with regard to their safety. Use of thoughtful risk‐based testing strategies can conservatively mitigate risk, allowing more of these devices to reach full maturity in development and arrive on the market.
References
1 1 Rebelo, R., Fernandes, M., and Fangueiro, R. (2017). Biopolymers in medical implants: a brief review. Procedia Engineering 200: 236–243.
2 2 Yadav, P., Yadav, H., Shah, V.G. et al. (2015). Biomedical biopolymers, their origin and evolution in biomedical sciences: a systematic review. Journal of Clinical and Diagnostic Research: JCDR 9 (9): ZE21.
3 3 Sahana, T. and Rekha, P. (2018). Biopolymers: applications in wound healing and skin tissue engineering. Molecular Biology Reports 45: 2857–2867.
4 4 Labet, M. and Thielemans, W. (2009). Synthesis of polycaprolactone: a review. Chemical Society Reviews 38 (12): 3484–3504. https://doi.org/10.1039/B820162P.
5 5 Chien, L.J. and Lee, C.K. (2007). Enhanced hyaluronic acid production in Bacillus subtilis by coexpressing bacterial hemoglobin. Biotechnology Progress 23 (5): 1017–1022.
6 6 Chien, L.J. and Lee, C.K. (2007). Hyaluronic acid production by recombinant Lactococcus lactis. Applied Microbiology and Biotechnology 77 (2): 339–346.
7 7 Liu, L., Wang, M., Du, G., and Chen, J. (2008). Enhanced hyaluronic acid production of Streptococcus zooepidemicus by an intermittent alkaline‐stress strategy. Letters in Applied Microbiology 46 (3): 383–388.
8 8 Helenius, G., Bäckdahl, H., Bodin, A. et al. (2006). In vivo biocompatibility of bacterial cellulose. Journal of Biomedical Materials Research Part A 76A (2): 431–438.
9 9 Hoenich, N.A. (2007). Cellulose for medical applications: past, present, and future. BioResources 1 (2): 270–280.
10 10 Rockwood, D.N., Preda, R.C., Yücel, T. et al. (2011). Materials fabrication from Bombyx mori silk fibroin. Nature Protocols 6 (10): 1612–1631.
11 11 Auras, R.A., Lim, L.‐T., Selke, S.E., and Tsuji, H. (2011). Poly(lactic Acid): Synthesis, Structures, Properties, Processing, and Applications. Wiley.
12 12 ISO/TS 21726:2019 (2016). Biological evaluation of medical devices – Application of the threshold of toxicological concern (TTC) for assessing biocompatibility of medical device constituents. Standardization IOf.
13 13 Kroes, R., Kleiner, J., and Renwick, A. (2005). The threshold of toxicological concern concept in risk assessment. Toxicological Sciences 86 (2): 226–230.
14 14 Munro, I., Renwick, A., and Danielewska‐Nikiel, B. (2008). The threshold of toxicological concern (TTC) in risk assessment. Toxicology Letters 180 (2): 151–156.
15 15 Hennes, E. (2012). An overview of values for the threshold of toxicological concern. Toxicology Letters 211 (3): 296–303.
16 16 Guideline IHT (2014). Assessment and control of DNA reactive (mutagenic) impurities in pharmaceuticals to limit potential carcinogenic risk. International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH), Geneva. Citeseer (31 May 2017). http://academy.gmp-compliance.org/guidemgr/files/M7_R1_ADDENDUM_STEP_4_2017_0331.PDF
17 17 European Food Safety Authority and World Health Organization (2016). Review of the Threshold of Toxicological Concern (TTC) Approach and Development of New TTC Decision Tree. EFSA Supporting publication 2016:EN‐1006. https://efsa.onlinelibrary.wiley.com/doi/pdf/10.2903/sp.efsa.2016.EN-1006
18 18 Cramer, G., Ford, R., and Hall, R. (1976). СКАЧАТЬ